The mouthpiece type factor: impact of filter selection on secondary electrospray ionisation high-resolution mass spectrometry (SESI-HRMS) breath profiles

R. Yassin, D. M. Baur, N. A. Sievi, K. Fricke, X. Zhang, S. Bajrami, K. D. Singh, F. Schmidt, M. Kohler

Abstract

Standardising Sampling for Reproducible Breathomics Research

Using EXHALION and Super SESI coupled to an Orbitrap mass spectrometer, this study evaluated seven mouthpiece types across 143 valid SESI-HRMS breath measurements. Mouthpiece selection accounted for 42.7% of molecular feature variance, while the unfiltered reference detected 397 features not consistently observed with filtered devices. Statistical correction reduced systematic variability but could not recover missing signals, highlighting the importance of standardised sampling for reproducible breathomics research.

View on the original journal

Previous
Previous

Exhaled breath metabolomics: Clinical diagnostic applications, technological evolution and translational challenges

Next
Next

Non-Invasive Estimation of Systemic Drug Exposure Using Exhaled Breath Metabolomics in Humans: A Translational Proof-of-Concept Study